OBJECTIVE
This SOP describes the procedure for the Calibration of Cold Chambers / Freezers.
SCOPE
To establish a Standard Operating Procedure to Calibrate the COLD CHAMBERS / FREEZERS
RESPONSIBILITIES
The Engineering Personnel is responsible for ensuring adherence to this procedure.
It is the Engineering Manager's responsibility to follow this procedure's instructions.
MATERIAL & EQUIPMENT
- Calibrated thermometer.
- Temperature bath.
- Computer with Microsoft Excel for error & uncertainty calculation
TYPE OF TEMPERATURE MONITORING SYSTEM
- Digital Temperature monitoring system
- Analog/Manual monitoring system
Based on the monitoring system followed by the organization, one of the following calibration methods should be adopted.
Method 1: (Calibration of Digital Temperature monitoring system)
PRELIMINARY INSPECTION
Before proceeding with calibration, the under-test temperature module should inspect the following:
- Remove the temperature module from the instrument & place a calibrated thermometer for temporary monitoring.
- Ensure all previous data are saved in the system.
- Check & ensure that the battery level is full
- Clean the sensor with 70% alcohol.
PROCEDURE
- Calibration shall be performed in the controlled calibration room as the ambient temperature 22°C ± 2°C and humidity between 50% to 75% is maintained.
- Expose both under test Thermometer as well as the Reference Thermometer to ambient air for 15 minutes.
- Switch on the temperature bath & wait for 15 minutes to stabilize the temperature.
- Place both thermometers in the temperature bath & wait for 15 minutes to stabilize the temperature.
- Let the temperature of the under-test thermometer (module) stabilize for one point.
- The reading of the under-test Thermometer as well as the Reference thermometer should be taken.
- Take the average of 5 readings of 3 different temperatures in the normal operating range & the data shall be filled in the excel format.
- Calculate the error taking the master’s error into account
- If the difference is > ± 1°C, adjust the temperature of under test Thermometer (temperature module).
- Calculate the Expanded Uncertainty & combined uncertainty of all measuring parameters, the uncertainty of the master certificate & least count of UUC. Providing a level of confidence of approx. 95%.
- Affix the calibration sticker on the temperature module.
- Clear the memory of the temperature module & place it back into the respective equipment.
- Make an entry in the service record, update the label affixed on the respective equipment
- Prepare a calibration report which should be documented in the engineering department.
Method 2: (Calibration of Analog/Manual monitoring system)
Before proceeding with calibration, the under-test thermometer is inspected for the following:
- Remove the thermometer from the instrument & place another calibrated thermometer for temporary monitoring.
- Cleaned & scales are visible.
- Entrapped Gases - in the bulb, contraction chamber and expansion chamber.
- Detached Mercury/Red Alcohol - at the top of the contraction chamber, expansion chamber and in the stem.
In the case one or both of the above faults are observed in the thermometer, such defects are first rectified.
Take reference thermometer and the under-test thermometer are exposed to the ambient air for 30 minutes kept in the vertical position. Then they are heated to their maximum nominal temperature for 20 minutes. Then both the thermometers are allowed to rest in the ambient air before starting calibration.
PROCEDURE
- Ensure the ambient temperature 22°C ± 2°C is maintained.
- Expose both under test Thermometer as well as the Reference Thermometer to ambient air for 15 minutes.
- Switch on the temperature bath & wait for 15 minutes to stabilize the temperature.
- Place both thermometers in the temperature bath & wait for 15 minutes to stabilize the temperature.
- Let the temperature of the under-test thermometer (module) stabilize for one point.
- The reading of the under-test Thermometer as well as the Reference Thermometer should be taken.
- Take the average of 5 readings of 3 different temperatures in the normal operating range & the data shall be filled in the excel format.
- Calculate the error taking the master’s error into account
- If the difference is > ± 1°C, replace the Thermometer and repeat the calibration.
- Calculate the Expanded Uncertainty & combined uncertainty of all measuring parameters, the uncertainty of the master certificate & least count of UUC. Providing a level of confidence of approx. 95%.
- Make an entry in the service record, update the label affixed on the respective equipment
- Prepare a calibration report which should be documented in the engineering department.
ABBREVIATION
QC: Quality Control
ANNEXURE
Nil
REVISION HISTORY
Nil
0 Comments