SOP for Performance Check of High-Performance Thin Layer Chromatography

Objective
To check the performance of High-performance thin layer chromatography.

Scope
This procedure is applicable to calibrate high-performance thin layer chromatography.

Responsibility
Doing: Technical Assistant
Checking: Executive /Manager

Accountability
Head of the Department

Frequency
once in a 3 month

Procedure
  • Check the performance of the instrument in terms of
  1. Linearity of spotting ( Annexure-I)
  2. Reproducibility of spotting ( Annexure-II)
  • Prepare the mobile phase as per the method given in Annexure-I.
  • Prepare the solution as mentioned in Annexure-I
  • Use 10 x 10 cm HPTLC plate (Kieselgel 60F254 or Identical)
  • For Linearity of spotting
  1. Apply 2 µl, 4µl, 6µl, 8µl and 10 µl of solution on the HPTLC plate with a spotter.
  2. Allow the plate to run in the mobile phase.
  3. Dry the plate with a drier.
  4. Scan the plate with the scanner.
  5. Check the linearity and correlation coefficient.
  6. Fill in the Data as mentioned in Annexure-I
  • Reproducibility of spotting
  1. Apply 10µl solution on the HPTLC plate for five times in sequence
  2. Allow the plate to run in the mobile phase.
  3. Dry the plate with a drier.
  4. Scan the plate with the scanner.
  5. Calculate the RSD for five tracks ( RSD Limit: N.M.T 3.0%)
  6. Fill in the data mentioned in Annexure-II

Abbreviations
RSD= Relative standard deviation
N.M.T = Not more than
N.L.T = Not less than
HPTLC = High-performance thin layer chromatography
API= Active pharmaceuticals ingredient


Annexure

Revision History
Nil