OBJECTIVE
To lay down the procedure for operation and calibration of the Disintegration Test Apparatus for suppositories for better and error-free use.
SCOPE
This procedure is applicable for the Disintegration Test Apparatus for suppositories installed in the Quality Control Department.
RESPONSIBILITY
Execution: Chemist and above – QC department.
Checking: Asst. Manager and above – QC department
OPERATIONAL PROCEDURE
- Fill the acrylic tank with distilled water up to the water level mark. Keep the safety acrylic lid on the tank.
- Connect the instrument to Mains and put the MAIN switch ‘ON’. The red pilot lamp and four-line LCD display will come on.
- Set the temperature of the medium and allow it to stabilize in the range of 36°C to 37°C.
- Introduce one tablet, or capsule into each tube and if prescribed in the product’s monograph, add a disk to each tube.
- Suspend the assembly in the beaker and reset the timer to ‘00’, then switch on the motor.
- After the specified time switch off the motor and mains and remove the assembly from the liquid.
CALIBRATION PROCEDURE
Frequency: Once in 3 months
- Operate the instrument as described above in the operation procedure.
- Temperature Setting: When the indicator displays 37°C, measure the temperature of the bath with a thermometer and record the results as per Annexure- 1
ABBREVIATION
SOP - Standard Operating Procedure
Q.C - Quality control
°C - Degree centigrade
ANNEXURE
Annexure-1: Calibration Record of Disintegration Test Apparatus.
REVISION HISTORY
Nil