SOP for Operation Procedure for Melting Point Apparatus

OBJECTIVE
To lay down the procedure for the Operation of the Melting Point Apparatus for better and error-free use

SCOPE
This procedure is applicable for the Melting Point Apparatus, installed in the Quality Control Department.


RESPONSIBILITY
Execution: Chemist and above - QC department.
Checking: Asst. Manager and above - QC department.


PROCEDURE
  • Fill the capillaries with the sample powder for melting point determination or mini test tubes with sample liquid in case of boiling point determination.
  • Insert the capillaries / mini test tubes through their appropriate openings on the right-hand side of the vessel.
  • Switch on the power supply and illumination.
  • Adjust the stirrer speed to its lowest and increase it gradually.


  • Adjust the heating control gradually to get a rate of increase of 30C / minute initially and 10C/minute in the vicinity of the expected melting/boiling point.
  • Observe the substance in the capillaries / mini test tubes through the magnifier.
  • When the substance starts melting or bubbles start forming, push the ‘DISPLAY HOLD SWITCH’ and note down the reading on the digital display.
  • Switch OFF the apparatus and allow it to cool.

ABBREVIATION
SOP: Standard Operating Procedure
Q.A: Quality Assurance
No. : Number
Q.C: Quality control
Dept. : Department

REVISION HISTORY
Nil