OBJECTIVE
To lay down the procedure for cleaning top Loading balances.
SCOPE
This SOP shall be applicable for all Top-loading balances, installed in the Quality Control Laboratory at XYZ Limited
RESPONSIBILITY
Execution: Chemist and above-QC department
PROCEDURE
Frequency: Daily
- Switch “OFF” the Instrument.
- Switch “OFF” the Mains.
- Remove the dust and powder with the help of a soft brush. Soak the tissue paper in Isopropyl alcohol and squeeze out the Isopropyl alcohol and clean the balance to remove oily substances or any spots inside the balance.
- Finally, clean it with tissue paper and let it dry in the air. Care is to be taken for proper drying when using isopropyl alcohol.
- Clean the outer surface with tissue paper wetted with Isopropyl alcohol then clean with dry tissue paper.
- After cleaning and drying the Top Loading balances, check the balance level.
- After cleaning, restart the balances and recalibrate it as per the current version of SOP.
- After cleaning the balances, record the cleaning details in the instrument log card as per the current version of SOP.
ABBREVIATION
SOP: Standard Operating Procedure
Q.C: Quality control
No: Number
ANNEXURE
Nil
REVISION CARD
Nil