SOP for Calibration Procedure for Hygro-Thermometer

OBJECTIVE
To lay down a procedure for calibration of the hygro-thermometer.

SCOPE
This procedure applies to the production department in the Pharmaceutical Manufacturing Facilities of XYZ Limited.


RESPONSIBILITY
  • The IPQA staff is responsible for the implementation of the process.
  • The QA In-Charge is responsible for the execution and compliance of the procedure.



PROCEDURE
  • Measuring Range: Temperature – 10℃ to 50℃
  • RH – 20% to 95%.
  • Collect in-house hygro-thermometer and compare it with standard hygro-thermometer.
  • Standard hygro-thermometer should be validated by the external body.
  • Check the date of validity of the standard hygro-thermometer.
  • Keep in-house hygro-thermometer with standard hygro-thermometer in one place.
  • Check the reading room in-house and standard hygro-thermometer at 9.00AM, 1.00PM, and 5.00PM and record it in the calibration log.
  • The average tolerance limit is +/-5%R.H.
  • Frequency of calibration: Once a Month.
  • Calibration details shall be entered in Annexure-1.

ABBREVIATION
Q.A: Quality Assurance
SOP: Standard Operating Procedure
No. : Number


ANNEXURE
Annexure No. 1 - Calibration log for hygro-thermometer

REVISION HISTORY
Nil