OBJECTIVE
To provide a procedure for the calibration of the microcontroller melting point apparatus.
SCOPE
Applicable to the microcontroller melting point apparatus in QC Chemistry.
RESPONSIBILITY
- The task shall be performed by QC Executive.
- It is the responsibility of the Manager QC to ensure that this procedure is followed accordingly.
PROCEDURE
- Set up the instrument as per the standard operating procedure.
- Use the following reference standards to calibrate the instrument, and record the observation in Attachment-I.
- Vanillin
- Acetanilide
- Sulphanilamide
- Caffeine
- The melting point of the reference standards should be within the standard range
- The standard melting point range shall be taken from the COA of the standard.
- Calibration should be carried out once every six months or after maintenance of the instrument.
ATTACHMENT
Attachment 1: Calibration record of Micro Controller Melting Point Apparatus
ABBREVIATION
QC- Quality Control
COA - Certificate of Analysis
REVISION HISTORY
Nil