OBJECTIVE
To provide a procedure for the Operation & Maintenance of the Lovibond Tintometer.
SCOPE
To establish a procedure for an operation of the Lovibond Tintometer.
RESPONSIBILITY
QC Designee: Responsible for the operation of the instrument.
Section In-charge: Responsible for reviewing the procedure
QC Head: Responsible for implementing the procedure.
QA Head: Overall compliance.
Operation
- Plug in the Lovibond Benchtop illumination system and switch it on and wait for 10 minutes to warm-up the lamp.
- Fix the Lovibond Nessler 2150 and insert the colour standard disc chosen among NSH, NSB, and NSX based on the color comparison.
- Fill one Nessler cylinder to just above the 50 ml mark with the solution under examination and fit the anti-meniscus plunger.
- Stand the cylinder in the right-hand compartment of the Nessleriser.
- Fill another in the same way with deionized water and place in the left-hand compartment
- Rotate the disc until the closet color match is obtained, if necessary estimate between the nearest values.
- Record the value displayed as APHA or Hazen units
Use Lovibond Nessleriser 2250 for Disc CAA & CAB
- Fill one Nessler cylinder to just above the 250 mm. mark with the solution under examination and fit the anti-meniscus plunger.
- Stand the cylinder in the right-hand compartment of the Nessleriser
- Fill another in the same way with deionized water and place it in the left-hand compartment.
- Rotate the disc until the closet color match is obtained, if necessary estimate between the nearest values.
- Record the value displayed as APHA or Hazen units
For Disc 4/28 and 4/28A Use Lovibond Daylight 2000+ Comparator
- Fill a 40 mm. cell to the 20ml line with the sample and place it in the right-hand compartment of the comparator. Leave the left-hand compartment empty.
- Rotate the disc until the closet colour match is obtained, if necessary estimate between the nearest values.
- Record the value displayed as APHA or Hazen units
Maintenance
- To avoid excessive heating, do not leave the lamp turned on for longer than the duration of the measurement.
- Always avoid touching the lamp as this can significantly reduce the life of the lamp
- The glass filter can be cleaned using soapy water and then rinsed.
- Clean the cells every time after use
Frequency: As and when required.