OBJECTIVE
To provide a procedure for the operation of the COD Digester.
SCOPE
This procedure applies to COD Digester at XYZ Ltd.
RESPONSIBILITY
Quality Control Chemist and above
ACCOUNTABILITY
Quality Control Head
PROCEDURE
Precaution
- Ensure that the instrument is clean and free from dust.
- Keep the instrument 1 foot away from the wall.
- Keep the instrument away from the washbasin or any water source.
- Ensure that any solvent/chemical does not fall on the aluminum block when it is hot.
- Use asbestos gloves while handling the hot tubes.
- Do not try to remove the tubes from the COD block and keep it in the tube stand.
- Do not try to remove the condenser tube when it is hot.
- Switch OFF the instrument when not in use.
- Calibrate the digital timer controller and heating sensor of the COD Digester once a year by the approved outside agency.
Cleaning
- Disconnect the instrument from the electrical connection.
- Clean the instrument with a soft cloth.
- Clean the tubes as per SOP of Cleaning of Laboratory Glassware.
Operation
- Connect the instrument to the main supply. Switch on the main supply.
- The red-colored bulb will glow indicating the power supply is connected.
- The orange-colored bulb will glow indicating the heater is on.
- The orange bulb remains on until the temperature is attained and will be switched off when the desired temperature is attained.
- Adjust the required temperature and time with the help of keys given on the digital control panel.
- Press the SET key.
- The display will show SP for temperature.
- Set 150°C temperature with the help of the up and down keys.
- Again press the SET key.
- The display will show soak time.
- Set 2.0 Hours with the help of up and down keys.
- In the soak time parameter, the digits before the decimal are hours & after the decimal it is minutes. i.e, (hh: mm)
- Fill 30 ml of the content of the sample in each tube and tightly close it with the condenser tube.
- Place 30 ml filled tubes in the aluminum COD block of the COD Digester for 2 hours.
- After completing the cycle, a digital alarm for 10 seconds will be produced and the heating will get automatically stopped.
- Allow cooling at room temperature.
- Perform further analysis as per the Specification of Chemical Oxygen Demand.
ABBREVIATIONS
COD: Chemical oxygen demand
SP: Set point
°C: Degree Celsius
hh: mm
Hours: Minutes
%: Percentage