SOP for Calibration of Glass Thermometer

OBJECTIVE
To provide a standard operating procedure for the calibration of Mercury mercury-filled glass thermometer.

SCOPE
This procedure applies to all thermometers used in plants.

RESPONSIBILITY
Doing: Technical Assistant/Executive
Checking: Executive /Manager

ACCOUNTABILITY
Head of the Department

PROCEDURE

Frequency: Once in a six month
  • Assign a Q.C Report No to each thermometer as per the following logic XXX/0000 
Where
XXX = Abbreviation of Company
0000 = serial number.

  • Keep the thermometer which is to be calibrated at room temperature till room temperature is achieved.
  • Visually inspect the thermometer for cracks, gaps, clear correct markings, chipped or distorted glass, and separation in the mercury column.
  • If separation in the mercury (liquid) column, is found then take the following corrective action.
  1. Immerse the bulb portion of the thermometer in a block of dry ice until the meniscus has stopped moving. Ensure not to freeze the mercury (liquid) in the bulb.
  2. If the separation remains, gently tap the bulb on a soft object and swing the thermometer in a small arch until the mercury (liquid) is united.
  3. Thermometers with measurement ranges above 50℃ are calibrated by a third party.
  • Calibration Procedure for thermometer having measuring range up to 50℃.
  1. First, keep all 50 range thermometers at room temperature till room temperature is achieved.
  2. After achieving room temperature, keep all thermometers in an ice bath with reference thermometers.
  3. Ensure the thermometer is immersed to the proper immersion depth.
  4. When the meniscus has stopped moving gently tap the thermometer, as some amount of the meniscus may adhere to the wall of the thermometer.
  5. View the 0 mark and compare the test thermometer with the reference thermometer.
  6. Record the value of the test and reference thermometer in Annexure-I.
  7. Use a low-temperature water bath for calibration.
  • For the first temperature point use the following procedure.
  1. Set the water bath at 10℃ ± 1 temperature. After achieving the temperature of 10℃ ± 1, keep the reference calibrated thermometer and test thermometer inside the water bath.
  2. After 5 minutes observe the temperature of the reference and test thermometer.
  3. Record the value of the test and reference thermometer in Annexure-I.
  • For the second temperature point use the following procedure.
  1. Set the water bath at 25 ± 1. After achieving the temperature of 25 ± 1, keep the reference calibrated thermometer and test thermometer inside the water bath.
  2. After 5 minutes observe the temperature of the reference and test the thermometer.
  3. Record the value of the test and reference thermometer in Annexure-I.
  • For the third temperature point use the following procedure.
  1. Set the water bath at 40 ± 1. After achieving the temperature of 40 ± 1, keep the reference calibrated thermometer and test thermometer inside the water bath.
  2. After 5 minutes observe the temperature of the reference and test the thermometer.
  3. Record the value of the test and reference thermometer in Annexure-I.
  • Ensure the results are within the tolerance limit.
  • If values are out of tolerance, keep the thermometer for 72 hrs. For stabilization and then recalibrate it . Remove it from the services and destroy the thermometer if it is out of the limit.
  • Maintain the issue record of the thermometer as per Annexure II.

ANNEXURE
Annexure-I: THERMOMETER CALIBRATION RECORDS
Annexure-II: THERMOMETER ISSUE RECORDS



ABBREVIATIONS
℃: Degree centigrade

REVISION HISTORY
Nil

Post a Comment

0 Comments

Close Menu
close