OBJECTIVE
To lay down a procedure for the operation of pass boxes.
SCOPE
This SOP applies to all Pass Boxes located in the Microbiology Laboratory of PharmaInfo Limited
RESPONSIBILITY
Microbiologist – Responsible for operating.
Section In-charge – Responsible for reviewing the procedure.
QC Head – Responsible for implementing the procedure.
QA Head – Overall compliance.
PROCEDURE
Operation
- Disinfect the outer and inner surfaces of the pass box with 70% IPA.
- Switch on the main supply.
- Switch ON the blower 30 minutes before the operation.
- Switch ON the UV light for 30 minutes before operation.
- Load the sterilized items i.e., microbial media flasks, sampling bottles, etc.
- Ensure that the pass box doors are in closed condition, after loading.
- Switch on the UV for 20 minutes to ensure the surface sterilization of loaded sterilized items.
- Record the pressure gauge reading and UV lamp burning hours in Annexure I.
- Operation of the pass box shall be followed carefully, and precautions like the opening of both doors at the same time shall be avoided.
- Every day after the completion of work pass box shall be cleaned using 70% Isopropyl alcohol.
- If any spillage occurs during transferring or unloading of material it shall be immediately wiped off with 70% IPA.
- Disinfect the outer and inner surfaces of the pass box with 70% IPA.
- Switch on the main supply.
- Open the door by pressing the “Open” button.
- Load the pass box with tested plates, flasks, tubes etc.
- Ensure that the pass box doors are in closed condition, after loading.
- The UV light of the Pass box will always be in ON condition when the pass box is in closed condition.
- The blower and normal light will ON when the pass box is in open condition.
- Operation of the pass box shall be followed carefully, and precautions like opening both doors at the same time shall be avoided.
- Every day after the completion of work pass box shall be cleaned using 70% Isopropyl alcohol.
- If any spillage occurs during transferring or unloading of material it shall be immediately wiped off with 70% IPA.
- Disinfect the outer and inner surfaces of the pass box with 70% IPA.
- Switch on the main supply.
- Open the door by pressing the “Open” button.
- Load the pass box with destructed media.
- Ensure that the pass box doors are in closed condition, after loading.
- The blower and normal light will be in ON condition when the pass box is in open condition.
- Operation of the pass box shall be followed carefully, and precautions like the opening of both doors at the same time shall be avoided.
- Every day after the completion of work pass box shall be cleaned using 70% Isopropyl alcohol.
- If any spillage occurs during transferring or unloading of material it shall be immediately wiped off with 70% IPA.
- Sanitize all Pass boxes with disinfectant as per the current version of the SOP Sanitization procedure for microbiology equipment and record the sanitization details in the respective log.
Calibration
- Calibration of the Pass box shall be carried out by an external agency once in year ± 30 days or after the maintenance job.
- UV lamps shall be replaced once in a year or after 9000 thousand burning hours which is earlier.
ABBREVIATIONS
QA: Quality Assurance
QC: Quality Control
SOP: Standard Operating Procedure
%: Percentage
IPA: Iso Propyl Alcohol
UV: Ultra Violet