OBJECTIVE
To lay down the procedure for cleaning analytical balances.
SCOPE
This SOP shall be applicable for all Analytical Balances, installed in the Quality Control Laboratory.
RESPONSIBILITY
Execution: Chemist and above-QC department
PROCEDURE
Frequency: Daily
- Switch “OFF” the Instrument.
- Switch “OFF” the Mains.
- Open the sliding doors of the balance and remove the pan which is inside the balance.
- Remove the dust and powder with the help of a soft brush. Soak the tissue paper in Isopropyl alcohol and squeeze out the Isopropyl alcohol and clean the balance to remove oily substances or any spots inside the balance.
- Finally, clean it with tissue paper and let it dry in the air. Care is to be taken for proper drying during the use of isopropyl alcohol.
- Clean the outer surface with tissue paper wetted with Isopropyl alcohol then clean with dry tissue paper.
- Also, clean the sliding doors of analytical balances with Isopropyl alcohol than with dry tissue paper.
- After cleaning and drying, close the sliding doors of the analytical balances and check the level of balances.
- After cleaning, restart the balances and recalibrate it as per the current version of SOP.
- After completion of cleaning of the balances, record the cleaning details in the instrument log card as per the current version of SOP.
- Place the ‘self-indicating’ coarse silica gel inside the balance.
- If the color of the silica gel changes from blue to pink, regenerate the silica gel by keeping it in the oven for some time till it gets blue color again.
REFERENCE
Operation and calibration procedure for analytical balances
Use of Instrument Log Card for Instrument / Equipment Status
ABBREVIATION
SOP - Standard Operating Procedure
Q.C - Quality control
No - Number
ANNEXURE
Nil
REVISION HISTORY
Nil
0 Comments