SOP for Calibration of High Performance Thin Layer Chromatography (HPTLC)

OBJECTIVE
To check the performance of High-performance thin-layer chromatography.

SCOPE
This procedure is applicable to calibrate high-performance thin-layer chromatography.

RESPONSIBILITY
Doing : Technical Assistant
Checking : Executive /Manager

ACCOUNTABILITY
Head of the Department


PROCEDURE
  • Check the performance of the instrument in terms of
  1. Linearity of spotting (Annexure-I)
  2. Reproducibility of spotting (Annexure-II)
  • Prepare the mobile phase as per the method given in Annexure-I.
  • Prepare the solution as mentioned in Annexure-I
  • Use 10 X 10 cm HPTLC plate (Kieselgel 60F254 or Identical)


For Linearity of Spotting
  1. Apply 2 μl, 4μl, 6μl, 8μl and 10 μl of solution on the HPTLC plate with a spotter.
  2. Allow the plate to run in the mobile phase.
  3. Dry the plate with a drier.
  4. Scan the plate with the scanner.
  5. Check the linearity and correlation coefficient.
  6. Fill in the Data as mentioned in Annexure-I

Reproducibility of Spotting
  • Apply 10μl solution on the HPTLC plate for five times in sequence
  • Allow the plate to run in the mobile phase.
  • Dry the plate with a drier.
  • Scan the plate with the scanner.
  • Calculate the RSD for five tracks ( RSD Limit: N.M.T 3.0%)
  • Fill in the data mentioned in Annexure-II


ABBREVIATIONS
%= Percentage
RSD= Relative standard deviation
N.M.T = Not more than
N.L.T = Not less than
μl = Microlitter
HPTLC = High-performance thin layer chromatography
API= Active pharmaceuticals ingredient

ANNEXURE

REVISION HISTORY
Nil