OBJECTIVE
To check the performance of High-performance thin-layer chromatography.
SCOPE
This procedure is applicable to calibrate high-performance thin-layer chromatography.
RESPONSIBILITY
Doing : Technical Assistant
Checking : Executive /Manager
ACCOUNTABILITY
Head of the Department
PROCEDURE
- Check the performance of the instrument in terms of
- Linearity of spotting (Annexure-I)
- Reproducibility of spotting (Annexure-II)
- Prepare the mobile phase as per the method given in Annexure-I.
- Prepare the solution as mentioned in Annexure-I
- Use 10 X 10 cm HPTLC plate (Kieselgel 60F254 or Identical)
For Linearity of Spotting- Apply 2 μl, 4μl, 6μl, 8μl and 10 μl of solution on the HPTLC plate with a spotter.
- Allow the plate to run in the mobile phase.
- Dry the plate with a drier.
- Scan the plate with the scanner.
- Check the linearity and correlation coefficient.
- Fill in the Data as mentioned in Annexure-I
Reproducibility of Spotting
- Apply 10μl solution on the HPTLC plate for five times in sequence
- Allow the plate to run in the mobile phase.
- Dry the plate with a drier.
- Scan the plate with the scanner.
- Calculate the RSD for five tracks ( RSD Limit: N.M.T 3.0%)
- Fill in the data mentioned in Annexure-II
ABBREVIATIONS
%= Percentage
RSD= Relative standard deviation
N.M.T = Not more than
N.L.T = Not less than
μl = Microlitter
HPTLC = High-performance thin layer chromatography
API= Active pharmaceuticals ingredient
ANNEXURE