SOP for Validation of Steam Sterilization

Purpose
It is established for the proper validation of the sterilization process.

Scope
It is applicable to microbiological Labs.

Responsibilities
  • Production Manager
  • Quality Control Manager
  • Microbiologist.


Procedure
  • Biological indicators to be used: Strain of bacillus stearothermophilus ATCC 7953Ø
  • Take 3-4 ampoules of the above-mentioned biological indicator in a container where the other filled ampoules are to be sterilized in an autoclave.
  • These ampoules of biological indicators are placed at different places in the autoclave to check the proper sterilization process. Close the access door and continue to operate the apparatus.
  • After completion of the sterilization procedure, and at a noted time within 2 hours, aseptically add each ampoule content in 30 ml of Soybean Casein Digest Medium.
  • Incubate each media at a temperature of 55-60°C. Observe each inoculated medium at 24 and 48 hours, and every 1 or 2 days after that for 7 days. If after 7 days of incubation, there is no growth in the medium the test for sterilization is met.

Abbreviations
SOP: Standard Operating Procedure. 

Annexure
Nil

Revision History
Nil

Post a Comment

0 Comments

Close Menu
close