Purpose
To describe a procedure for carrying out validation of LAF and to check the effectivity of UV light.
Scope
This SOP is applicable to the validation of Laminar Airflow.
Responsibility
Microbiologist or QC -In charge.
Procedure
Validate the laminar flow by following tests:
Particle count:
- This test method specifies the measurements of air-born particle concentration with size distributions.
- The system should be kept ON 2hrs before starting measurements.
- Ensure that the floor and area as well as equipment is visually cleaned.
- Use a calibrated particle counter and carry out particle counting at corners and centers of the Laminar Airflow unit.
- 0.5 μ NMT 100 particle/ft3 of air
- 5.0 μ NIL particle/ft3 of air
- Record the mean particle count in the format for “Report of Validation of Laminar Air Flow”.
Air Flow Velocity Measurement:
- This test is performed to determine the average airflow velocity, volume, and uniformity, to determine the air supply volume flow rate.
- Clean the instrument with IPA and the system should be kept ON before starting the measurements.
- Measure the air velocity near the HEPA filter at a distance of 6 inches below the filter.
- Carry out measurements by using a calibrated Anemometer. Check the velocity at the corners and center of the HEPA – Filter. The limit is 40-80 fpm.
DOP Test:
- DOP Test is performed by introducing DOP aerosol upstream of the HEPA – filter and searching for leaks by scanning the downstream of the HEPA filter with a calibrated photometer probe.
- Measure the upstream aerosol concentration using a calibrated photometer. Adjust tested and search the area for any leakage.
- DOP Test passes with penetration less than 0. 003 %.
- The report of DOP testing will be attached and received from the authorized agency.
Frequency: Particle Count: 6 months, Air Velocity: 12 Months, DOP Test: 24 Months.Note: DOP Test, Air Velocity Measurement, and particle count monitoring testing should be carried out by an authorized external agency.
Check the effectivity of the UV tube by the following procedure:
- Prepare Soyabean casein digest agar plates; incubate the plates for 24 hours at 30-35°C to check the sterility of the medium. Discard the contaminated plates, if any.
- Clean the surface of Laminar Air Flow with 70% IPA.
- Put ON the UV light for 30 minutes.
- Label the plates for the following:
- Name of the place
- Date of exposure
- After 30 min. put OFF UV light put ON normal light and Laminar Air Flow.
- Expose the Soya bean Casein Digest Agar plates for 1 hr.
- Incubate the plates at 35°C for 48-72 hrs.
- After completion of the incubation period count the cfu and record the results in the format “Verification of efficiency of UV”
- After incubation, cfu should be nil in all plates.
Frequency: Once a month.
Abbreviations
DOP : Dioctyl Pthalate
HEPA: High-Efficiency Particulate Filter
LAF: Laminar Air Flow
UV: Ultra Violet
CFU: Colony Forming Unit
FPM: Feet Per Minute
IPA: Iso Propyl Alcohol