SOP for Operation of Electrolab Dissolution Apparatus

OBJECTIVE
To lay down the Operation procedure of the dissolution test apparatus for better and error-free use.

SCOPE
This procedure is applicable for all dissolution test apparatus installed in the Quality Control Laboratory.

RESPONSIBILITY
Execution: Chemist and above – QC department.
Checking: Asst. Manager and above – QC department



PROCEDURE
  • Ensure that the working bench and instrument is clean.
  • Ensure that the instrument is calibrated.
  • Switch on the mains power.
  • Fill the bath with water up to the mark level.
  • Switch on the instrument.
  • When the power is on, the display shows the previously set time, RPM and the temperature display shows the bath temperature.
  • Place the lift jar filled with the medium according to the STP.
  • Fix the paddle/basket according to the product STP.
  • Press the temperature set switch and rotate the setting knob till the digital display shows 37°C.
  • Change the mode switch to RPM mode and switch on the motor.
  • Using a speed knob, adjust the RPM to the required level and switch off the motor.
  • Press both the Alarm and Reset switch so that the Timer display shows 0.00.
  • By pressing both Alarm and Slow or Alarm and Fast, set Alarm to the required time
  • Allow the bath temperature instrument.
  • After the test is completed the instrument will stop which is indicated by a beep sound.
  • Press the ↑ and ↓ keys to lift the stirrer unit.
  • Take the required amount of the sample from the jars and clean it.
  • Switch OFF the instrument.

ABBREVIATION
SOP - Standard Operating Procedure
Q.A - Quality Assurance
Q.C - Quality control
USP - United States Pharmacopoeia

ANNEXURE
Nil

REVISION HISTORY
Nil