Objective
To lay down a procedure for the operation and cleaning of the Laminar bench.
Responsibility
Microbiologist/Executive.
Accountability
Executive
Procedure
Introduction
- A laminar flow clean air workstation is a device that provides a total sweep of a confined area because the entire body of air moves with uniform velocity along parallel lines, originating through a HEPA filter occupying one entire side of the confined area. Therefore, it bathes the entire area with very clean air, sweeping away contaminants. Laminar flow provides well-controlled work areas only if proper precautions are observed.
Working
- Enter the sterility testing room/aseptic zone with proper gowning procedure.
- Clean the LAF with IPA 70% solution daily before operation and after the day's work.
- Check the electric connection before starting the work.
- Note the manometer reading. It should be zero if required and set at 'zero'.
- Switch ON the LAF unit and also the lights
- Check the manometer reading (It should be between 10-15min.)
- Switch ON the L.A.F. 1/2 hr. before sterility testing /aseptic operation.
- The Pre-filter should be cleaned every 3 months
Annexure
Nil
Revision History
Nil