Objective
The purpose of this SOP is to provide a written procedure for the operation of analytical balance.
Scope
This SOP covers operations of Analytical Balance, Make-Precisa, and Model-XB220A.
Responsibility
QC Officer (Analyst): Responsible for operation, daily calibration, and maintenance of the balance as per the procedure.
QA Officer/QA Manager: Review the records and govern the document.
Procedure
Ensure that the instrument is visibly clean and free from dust.
Setting and storing the configuration:
- Press << ON/OFF >> to switch on the balance.
- Continue to hold down the << MODE >> and the << T >> button during the start-up process, until the desired configuration appears on the display, and then releases the button. Select Store configuration as user configuration.
Working with the configuration menu:
- Press << ON/OFF >> to switch on the balance.
- Continue to hold down the << MODE >> button during the start-up process, until the desired configuration appears on the display, and then release the button. Set configuration as below.
- Language
- Unit - 1
- Set Data Print
- Set calibration
- Set Weighing Mode
- Set Interface
- Set Date and Time
- Password
- Theft code
1. Language
- Switch on the balance and depress << MODE >> until the currently activated language is shown.
- Release << MODE >> Press << MODE >> repeatedly until the language required is shown.
- Press << PRINT >> in order to confirm the selection.
- Press << PRINT >> in order to quit the menu.
2. Weight Unit
- Switch on the balance and depress << MODE >> until the currently activated language is shown.
- Release << MODE >>.
- Press << MODE >> repeatedly until the Weight unit is shown.
- Press << PRINT >> in order to confirm the selection.
- Press << PRINT >> in order to quit the menu.
3. Print Function
- Switch on the balance and depress << MODE >> until the currently activated language is shown.
- Release << MODE >>.
- Press << MODE >> repeatedly until the Set Data Print is shown.
- Set the required Parameters i.e. Auto Start ON/OFF, Mode, Time Base, Set Print Format, etc.
- Press << PRINT >> in order to confirm the selection. Press << PRINT >> in order to quit the menu.
4. Calibration Function
- Switch on the balance and depress << MODE >> until the currently activated language is shown.
- Release << MODE >>. Press << MODE >> repeatedly until the Set Calibration is shown.
- Set the required parameters. Press << PRINT >> in order to confirm the selection.
- Press << PRINT >> in order to quit the menu.
- Set all the configuration parameters by Switching on the balance and depressing << MODE >> until the currently activated language is shown. Release << MODE >>.
- Press << MODE >> repeatedly until the required parameter is shown. Set the required parameters.
- Press << PRINT >> in order to confirm the selection.
- Press << PRINT >> in order to quit the menu.
Setting and storing the application menu:
- Press << ON/OFF >> to switch on the balance.
- After the start-up process has finished, press the << MODE >> button, until the “SET APP. …” appears on the display. Set the required parameters i.e. Units, Count, Percent Balancing etc. Press << PRINT >> in order to confirm the selection. Press << PRINT >> in order to quit the menu.
- Press << MODE >> Button, until the “SET APPLICATION” appears on the display. Set the required parameters of selected under “Application”.
- Press << MODE >> Button, until the “AUTO-START ON/OFF” appears on the display.
The selected working program can, if required, be loaded automatically every time the balance is switched on.
Precaution
- The balance may only be used with the power adaptor.
- During weighing close the door properly.
- After weighing clean the pan with a hairbrush.
- Take care when weighing liquid is split into the inside of the balance or into connections on the rear of the equipment or the power adaptor.
Annexure
Annexure– 1 - Balance Usage log book.
Revision History
Nil