Ad Code

SOP for Operation and Calibration Procedure for Analytical Balance

OBJECTIVE 
To lay down the procedure for the operation and calibration of the Analytical Balance for better and error-free use.

SCOPE 
This procedure is applicable for Analytical Balances installed in the QC Department, IPQC Department & Dispensing Department.

RESPONSIBILITY 
It is the responsibility of QC Executives to follow the procedure without deviation. 

ACCOUNTABILITY 
Manager – Quality Control 



PROCEDURE 

Operation Procedure:
  • Make sure that the balance is kept clean & placed on a flat platform free from vibrations.
  • Ensure that the calibration status is valid.
  • Ensure that the spirit level is in the center of the circle.
  • Connect the power cable to the mains and switch ‘ON’.
  • Automatically self-checking starts & ends with OFF.
  • Press the ON/OFF key and all the displays will glow.
  • Press “TARE KEY”, and 0.0000 marks appear on the display.
  • The stability of the reading is obtained and is indicated by an arrow mark on the left side of the display.
  • Once the stability is attained, the balance is ready for weighing.
  • Place the material to be weighed on the pan & note down the reading after the arrow mark appears on the left side of the display.
  • Clean the balance immediately after weighing.

Preliminary Checks For Balance Calibration:

Spirit Level Check (Balance Leveling):
  • Adjust the air bubble to the center of the inner circle of the indicator by rotating the two leveling screws at the rear base of the balance.
  • Acceptance Criteria: The air bubble must be in the center of the inner circle.

Pan Cleaning and Adjustment:
  • Before placing the std. weight/butter paper/glassware, clean the pan properly. Also, ensure that the pan is adjusted properly (if not adjusted before).
  • Acceptance Criteria: Pan must be cleaned and centrally placed.

Zero Error Check:
  • Close the sliding doors of the balance press the Tare key and wait till the LCD shows zero.
  • Acceptance Criteria: LCD should display zero on the LCD screen.

Handling of Weight Box:
  • Ensure that the weights used are calibrated and certified on a regular basis.
  • Always use forceps/tweezers to place the weights on the balance pan.
  • Do not drop the weights with force on the pan of the balance, to avoid any variation in the results.
  • Weights used for calibration should always be cleaned with tissue paper before & after each use.
  • Acceptance Criteria: Weights should not have any denting or damage signs and should be cleaned before and after each use.

Calibration Procedure:

Frequency: 
  • Each working day: Spirit Level Check, Pan cleaning and adjustment, Zero Error Check, Accuracy.
  • Weekly & after maintenance: Spirit Level Check, Pan cleaning and adjustment, Zero Error Check, Accuracy, Precision/Measurement Uncertainty, and Corner accuracy/eccentricity test.
  • After preliminary checks are completed operate the instrument as per the operation procedure described above & proceed for external calibration.


Accuracy:
  • For daily calibration use 50 mg & 100mg. Place each weight on the weighing pan. Note down the observed weight.
  • For Weekly calibration use 50mg, 100mg, 200mg, 500mg, 1g, 5g, 10g, 50g, and 100g weights.
  • Record the readings as per Attachment - I for daily calibration and as per Attachment - II for Weekly calibration.
  • Acceptance Criteria: Each weighing should not differ by more than 0.1% of the actual weight.

Precision/ Measurement Uncertainty:
  • Place the 50mg weight on the weighing pan & note down the observed weight.
  • Repeat the above steps nine times.
  • Record the results as per Attachment - II
  • Calculate the Standard Deviation (SD).
  • Calculate the measurement uncertainty using the following equation:


  • Repeat the above steps for 100g weight, and record the results as per Attachment - II
  • Acceptance Criteria: Each weighing shall not differ more than 0.1% of the actual weight.
  • Measurement uncertainty should not be more than 0.001.

Corner Accuracy/Eccentricity Test:
  • Place 50g weight on the center, front-left, front-right, back-right and back-left side of the weighing pan (As shown below) and note down the weight for each position.
  1. Center
  2. Front - Left
  3. Front - Right
  4. Back - Right
  5. Back - Left 
  • Calculate the difference between the observed weight and the certified weight.
  • Record the results as per Attachment - II for Weekly calibration.
  • Acceptance Criteria: Each weighing shall not differ more than 0.1% of the actual weight.

ATTACHMENTS
Attachments I: Preliminary Checks & Daily Calibration Record of Analytical Balance 
Attachments II: Calibration Record of Analytical Balance


ABBREVIATIONS
QC – Quality Control
LCD – Liquid Crystal Display

REVISION HISTORY
Nil

Post a Comment

0 Comments

Close Menu
close